• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION GLADIATOR ELITE; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION GLADIATOR ELITE; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Model Number 24686
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/01/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Date of event: used the first day of the month of the bsc aware date since no event date provided.
 
Event or Problem Description
It was reported that balloon rupture occurred.A 10.0 x40, 75cm gladiator elite balloon catheter was advanced for dilatation.However, during inflation at 10 atmospheres, the balloon ruptured.No patient complications were reported.
 
Additional Manufacturer Narrative
B3 date of event: used the first day of the month of the bsc aware date since no event date provided.Device evaluated by mfr.: gladiator device 10x40x75cm was received for analysis.A visual examination identified that the balloon was not folded which indicates that the balloon was subjected to positive pressure.A microscopic examination identified a balloon longitudinal tear beginning approximately 9mm distal of the distal markerband and extending approximately 31mm proximally across the balloon material.An examination of the balloon material and markerbands identified no issues which could potentially have contributed to this complaint.The rated burst pressure for this device is 14 atmospheres.A visual and microscopic examination of the markerbands identified no issues which may potentially have contributed to the complaint incident.A visual and tactile examination found no kinks or damage to the shaft of the device.A visual and tactile examination identified no issues or damage to the tip of the device.No other issues were identified during the product analysis.
 
Event or Problem Description
It was reported that balloon rupture occurred.A 10.0 x40, 75cm gladiator elite balloon catheter was advanced for dilatation.However, during inflation at 10 atmospheres, the balloon ruptured.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLADIATOR ELITE
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key11334727
Report Number2134265-2021-01836
Device Sequence Number823898
Product Code LIT
UDI-Device Identifier08714729809517
UDI-Public08714729809517
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K132810
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/27/2022
Device Model Number24686
Device Catalogue Number24686
Device Lot Number0024029439
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/03/2021
Initial Date Received by Manufacturer 02/09/2021
Supplement Date Received by Manufacturer03/16/2021
Initial Report FDA Received Date02/16/2021
Supplement Report FDA Received Date03/22/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-