• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION SYNERGY; BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION SYNERGY; BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM Back to Search Results
Model Number 10605
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 01/29/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that stent dislodgement occurred.The target lesions were located in the left main (lm) artery and left circumflex (lcx) artery.The physician attempted to advance the first synergy stent to treat the lcx but the device could not cross the lm due to the tortuosity and heavy calcification and the physician decided to deploy the first stent in the lm.A 3.50 x 28mm synergy ii drug eluting stent was advanced to treat the distal lcx but it got caught on the first stent and the stent came off the balloon.An attempt was made to snare the dislodged stent but it was unsuccessful.The physician then advanced a small balloon through the undeployed stent and the stent was expanded where it was stuck.The procedure was successful.No patient complications were reported and the patient was not harmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY
Common Device Name
BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key11334837
Report Number2134265-2021-01425
Device Sequence Number823900
Product Code NIQ
UDI-Device Identifier08714729840305
UDI-Public08714729840305
Combination Product (Y/N)Y
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P150003
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/16/2022
Device Model Number10605
Device Catalogue Number10605
Device Lot Number0026418812
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/29/2021
Initial Report FDA Received Date02/16/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-