| Model Number |
VBH100502A |
| Health Effect - Clinical Codes |
Aneurysm (1708); Insufficient Information (4580)
|
| Date of Event |
08/14/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Cbas® heparin surface incorporates cbas-heparin manufactured from heparin sodium apl, which in covalently bound to the device surface and is essentially|non-eluting.
|
| |
|
Event or Problem Description
|
|
During treatment of a splenic aneurysm, a gore® viabahn® endoprosthesis (viabahn device) did not advance to the splenic artery due to tortuous anatomy.Attempts were made to withdraw the viabahn device through the 12fr cook sheath; however, the device could not be withdrawn.The device was deployed in the gastroduodenal artery and the deployment line was pulled out.However, the delivery catheter became stuck and could not be removed.Decision was made to have the transplant surgeon cut and remove the delivery catheter during a scheduled renal transplant surgery on (b)(6) 2020.(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the manufacturing records indicated the lot met all pre-release manufacturing specification.A product history review was conducted for the vbx device, and manufacturing records show the device lot met all pre-release specifications.This evaluation follows the complaint detail to confirm failure to advance and difficulty during withdrawal.Both are likely the result of tortuous anatomy; further causal identification is not possible.
|
| |
|