• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NIHON KOHDEN CORPORATION ZM-520PA; TRANSMITTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NIHON KOHDEN CORPORATION ZM-520PA; TRANSMITTER Back to Search Results
Model Number ZM-520PA
Medical Device Problem Codes Computer Software Problem (1112); Data Problem (3196); Patient Data Problem (3197)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/21/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The biomed reported that the previous day a patient had coded and they noticed data was not recorded.The data had about a 10-11min gap where it was blank.Waveforms were still displaying on the screen but not recorded.No known impact or consequence to the patient.Nihon kohden continues to investigate the reported event.The following fields are not applicable (na) to this report: b2 d4 lot # & expiration date d6a & d6b d7b d10 g1 - g8 h1- h9 the following fields contain no information (ni), as attempts to obtain information were made, but not provided: a2 - a6, b6, b7 attempt #1: 01/22/2021 emailed customer via microsoft outlook for patient information, relevant tests/labs and other relevant history: no reply was received.Attempt #2: 02/01/2021 emailed customer via microsoft outlook for patient information, relevant tests/labs, and other relevant history: no reply was received.Attempt #3: 02/08/2021 emailed customer via microsoft outlook for patient information, relevant tests/labs, and other relevant history: no reply was received.
 
Event or Problem Description
The biomed reported that the previous day a patient had coded and they noticed data was not recorded.The data had about a 10-11min gap where it was blank.Waveforms were still displaying on the screen but not recorded.No known impact or consequence to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZM-520PA
Common Device Name
TRANSMITTER
Manufacturer (Section D)
NIHON KOHDEN CORPORATION
attn: shama mooman
1-31-4 nishiochia
shinjuku-ku, tokyo 116-8 560
JA  116-8560
MDR Report Key11335638
Report Number2080783-2021-00083
Device Sequence Number19066254
Product Code DRT
UDI-Device Identifier04931921115077
UDI-Public4931921115077
Combination Product (Y/N)N
Initial Reporter StateMS
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Importer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 02/16/2021
Report Date (Section F) 02/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberZM-520PA
Device Catalogue NumberZM-520PA
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date01/22/2021
Device Age27 MO
Event Location Hospital
Date Report to Manufacturer02/16/2021
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date02/16/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Date Report Sent to FDA02/16/2021
-
-