Catalog Number 10562
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/22/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The reported defective device has been returned to the manufacturer and an investigation into the root cause for event is currently in progress.The results of the device evaluation will be sent via a follow up medwatch.Reference (b)(4).
 
Event or Problem Description
A radiology technician reported an issue with a exodus multipurpose 25 cm drainage catheter.During a check of the catheter, it was noted that the drain had lost suction and the hub was found to be cracked.The drain was removed and replaced with a new of the same drainage catheter.The patient did not experience any adverse effects, harm, or require medical intervention as a result of this incident.The reported device has been returned to the manufacturer and an investigation into the root cause for event is currently in progress.
 
Additional Manufacturer Narrative
Returned for evaluation was an exodus drainage catheter.As received, the drainage catheter hub was cracked.The drainage catheter was contained with bio material inside and out.Exodus drainage catheters are a purchased device from supplier/contract manufacturer freudenberg medical.Freudenberg has been notified via scar004263 for sample evaluation and dhr review of supplier lot.As per scar004263 results from freudenberg medical, visual review confirmed that the hub body was cracked for thereturned sample.A failure mode cannot be determined.A review has indicated that there are no issues with the molding process or the assembly process.These processes do not expose the hub to chemicals which could indue the crack.The device was manufactured in accordance with the dmr.This complaint is not manufacturing attributable.No correction is required as freudenberg medical did not attribute the hub crack to be manufacturing related; root cause cannot be determined.The customer's reported complaint description is confirmed, however the rott cause of the event could not be determined.A review of the device history records was performed for the reported packaging and component lots for any deviations related to the reported defect of the complaint.The review confirms that the lots met all material, assembly, and performance specifications.Labeling review: the instructions for use which is supplied to the end user with this catalog number states, "warnings: do not use this catheter with alcohol.Placing a 10 french or larger catheter as the primary drain before formation of a tract may be difficult in some patients.In these patients, the initial biliary drainage should be started with a smaller (8 french) catheter until a suitable tract allows placement of a larger catheter.Where long-term use is indicated, it is recommended that indwelling time not exceed 90 days.This catheter should be evaluated by the physician on or before 90 days post placement." labeling review: directions for use (dfu) aw1005142-01 states the following: caution: all connections must be secure and airtight.Perform inspections of the catheter and connections regularly.A review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Reference (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXODUS DRAINAGE CATHETER
Common Device Name
BILIARY (GENERAL) (NEPHOSTROMY) DRAINAGE CATHETER,,
Manufacturer (Section D)
ANGIODYNAMICS
10 glens falls technical park
glens falls, ny 12801 NY 12801
MDR Report Key11335679
Report Number1317056-2021-00023
Device Sequence Number16812200
Product Code FGE
UDI-Device IdentifierH965105621
UDI-PublicH965105621
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K152069
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2023
Device Catalogue Number10562
Device Lot Number0000154719
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/03/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/22/2021
Supplement Date Received by Manufacturer01/22/2021
Initial Report FDA Received Date02/16/2021
Supplement Report FDA Received Date04/19/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age73 YR
-
-