| Model Number |
OJR418 |
| Medical Device Problem Codes |
Use of Device Problem (1670); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The complaint ojr418 optiflow junior 2 nasal cannulas are currently en route to fisher & paykel healthcare (f&p) for evaluation.We will provide a follow-up report upon completion of our investigation.
|
| |
|
Event or Problem Description
|
|
A healthcare facility in (b)(6) reported, via a fisher & paykel healthcare (f&p) field representative, that the tubing of two ojr418 optiflow junior 2 nasal cannulas broke during use.It was noted that the tubing was possibly pulled or accidentally caught on equipment.There was no reported patient consequence.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).The optiflow junior 2 nasal cannula is designed specifically for the delicate anatomical features and flow requirements of neonatal and paediatric patients.It features adhesive pads (wigglepads) to maintain cannula stability on the patient's cheeks, soft-touch nasal prongs and breathable kink-proof and crush-resistant flexible tubing.Method: the two complaint ojr418 optiflow junior 2 nasal cannulas were not returned to fisher & paykel healthcare (f&p) for evaluation.Our investigation is thus based on the information provided by the customer and our knowledge of the product.Result: the customer reported that the tubing of two ojr418 optiflow junior 2 nasal cannulas broke during use.It was noted that the tubing was possibly pulled or accidentally caught on equipment.Conclusion: without the complaint devices, we are unable to determine the cause of the reported failure.All optiflow junior cannulas are 100% leak and occlusion tested after final assembly and any cannula that fails is discarded.Samples are also taken and pull tested to check the tube tensile strength and the glue joint strength at the cannula/tube joint, as well as the swivel grip joint.The subject cannula would have met the required specification at time of production.The user instructions illustrate in pictorial format the correct set-up and proper use of the ojr418 optiflow junior 2 nasal cannula.They also state the following: - appropriate patient monitoring (e.G.Oxygen saturation) must be used at all times.Failure to monitor the patient (e.G.In the event of an interruption to gas flow) may result in serious harm or death).- do not wrap, insulate, stretch or crush the tubing as this may impair the performance of this product or compromise safety (including potentially causing patient harm).- ensure all connections are secure during use.Check the cannula is undamaged and that the flow path is maintained.Under excessive load, the cannula may disconnect to prevent forces being transferred to the patient.
|
| |
|
Event or Problem Description
|
|
A healthcare facility in indiana reported, via a fisher & paykel healthcare (f&p) field representative, that the tubing of two ojr418 optiflow junior 2 nasal cannulas broke during use.It was noted that the tubing was possibly pulled or accidentally caught on equipment.There was no reported patient consequence.
|
| |
|