Model Number 20005D
Health Effect - Clinical Code Eye Injury (1845)
Date of Event 01/20/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A review of records related to the device including labeling, complaint trending, and risk documentation will be performed.Upon completion of this review, if there is any further relevant information obtained, a supplemental medwatch will be filed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported on (b)(6) 2021 that the ifs z-cal moved during the day and thickness was 130mm.The surgeon reported that a patient had a torn flap from 8 to 10 o'clock on one eye.The flap was put back down and covered with a bandage contact lens.
 
Additional Manufacturer Narrative
Correction: h4 - manufacturing date - the manufacturing site reported that the manufacturing date for the device is january 27, 2010.Additional information: h6 type of investigation codes 3331, 4110 and 4109.H6 investigation findings: code 213.Device evaluation: a review of the records related to this equipment that included labeling, manual, trending, and risk documentation reviews for this equipment were performed.Equipment labeling provides possible complications that can be caused by the surgical/treatment procedure being performed.The trend review shows that there is not a recognizable adverse trend.The risks and mitigations associated with the complaint issue are identified in existing risk documents and no new risks were identified as part of this investigation.There was no product deficiency identified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FS LASER SYSTEM
Common Device Name
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
AMO MANUFACTURING USA, LLC
510 cottonwood drive
milpitas CA 95035
MDR Report Key11337670
Report Number3006695864-2021-07180
Device Sequence Number4502482
Product Code GEX
UDI-Device Identifier05050474573451
UDI-Public(01)05050474573451
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
K063682
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/31/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number20005D
Device Catalogue Number20005D
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 01/20/2021
Supplement Date Received by Manufacturer03/04/2021
Initial Report FDA Received Date02/17/2021
Supplement Report FDA Received Date03/31/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-