MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE CONNECTED OR CART, 120 V; CARDIOPULMONARY EMERGENCY CART

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE CONNECTED OR CART, 120 V; CARDIOPULMONARY EMERGENCY CART Back to Search Results
Model Number 240099155
Medical Device Problem Code Loss of Power (1475)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/19/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Event or Problem Description
It was reported that there was loss of image.
 
Additional Manufacturer Narrative
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.The reported failure mode will be monitored for future reoccurrence.Alleged failure: hub lost power probable root cause: ¿ overloading the cart.¿ use error.Manufacturing date is unknown h3 other text : 81.
 
Event or Problem Description
It was reported that there was loss of image.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONNECTED OR CART, 120 V
Common Device Name
CARDIOPULMONARY EMERGENCY CART
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
MDR Report Key11338486
Report Number0002936485-2021-00092
Device Sequence Number19768846
Product Code BZN
UDI-Device Identifier00858701006360
UDI-Public00858701006360
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K082813
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number240099155
Device Catalogue Number0240099155
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 01/19/2021
Supplement Date Received by Manufacturer01/19/2021
Initial Report FDA Received Date02/17/2021
Supplement Report FDA Received Date04/11/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-