| Model Number |
IPN008819 |
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/25/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The device history review for the product visistat 35w 6/box, lot# 73f1900328.Investigation did not show issues related to the complaint.The device has not been returned for investigation.Teleflex will continue to monitor and trend related events.
|
| |
|
Event or Problem Description
|
|
It was reported that (b)(6) 2021, the staple was found jamming and could not be fired prior to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).The customer returned one unit 528235 visistat 35w 6/box for investigation.The returned sample was visually examined with and without magnification.Visual examination of the returned sample revealed that the staples appear misaligned.The sample was returned with its trigger partially engaged.The stapler appears used as there was biological material observed on the device.There was one partially loaded staple in the device.The stapler was returned with 31 staples left in the cover block indicating that at least 4 staples were fired by the end user.An attempt to fire staples was made by engaging the trigger of the stapler using hand pressure.Upon engagement of the trigger, the trigger cycle was completed , and no staple fired.Two more attempts were made, but no staples fired.It appeared that the misalignment of the staples was preventing the staples from moving down the track and into the forming tool.An attempt was made to manually realign the staples , but the staples were unable to be realigned.The stapler was disassembled , and no other defects or anomalies were observed with the device.It could not be determined what exactly caused the staples to misalign.A nonconformance has been opened by the manufacturing site to further investigate visistat jamming issues.It could not be determined what caused the staples to misalign.A nonconformance has been opened by the manufacturing site to further investigate this visistat jamming issue.The reported complaint of "misfire/jamming - staples not firing" was confirmed based upon the sample received.Visual examination revealed that the staples were misaligned.Upon functional inspection, the misalignment of the staples prevented any staples from firing.The sample was disassembled, and no other defects or anomalies were observed.It could not be determined what caused the staples to misalign.A nonconformance has been opened by the manufacturing site to further investigate visistat jamming issues.
|
| |
|
Event or Problem Description
|
|
It was reported that (b)(6) 2021, the staple was found jamming and could not be fired prior to the patient.
|
| |
|
Search Alerts/Recalls
|