Event or Problem Description
|
|
It was reported that balloon rupture occurred.The target lesion was extremely tight in a st-segment elevation myocardial infarction case.A 3.00mm x 15mm emerge balloon catheter was advanced for dilation.However, during procedure, the balloon ruptured.Another 3.00mm x 15mm nc emerge balloon catheter was advanced for dilation but it also ruptured during procedure.There were no patient complications nor injuries reported.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The target lesion was extremely tight in a st-segment elevation myocardial infarction case.A 3.00mm x 15mm emerge balloon catheter was advanced for dilation.However, during procedure, the balloon ruptured.Another 3.00mm x 15mm nc emerge balloon catheter was advanced for dilation but it also ruptured during procedure.There were no patient complications nor injuries reported.It was further reported that the target lesion was extremely tight and the balloons were inflated at 8 atmospheres.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The target lesion was extremely tight in a st-segment elevation myocardial infarction case.A 3.00mm x 15mm emerge balloon catheter was advanced for dilation.However, during procedure, the balloon ruptured.Another 3.00mm x 15mm nc emerge balloon catheter was advanced for dilation but it also ruptured during procedure.There were no patient complications nor injuries reported.It was further reported that the target lesion was extremely tight.The balloons were inflated at about 8 atmospheres.It was further reported that the patient had a critical ostial left anterior descending artery stenosis.
|
| |
|
Additional Manufacturer Narrative
|
|
(a3) sex - updated (b5) describe event or problem - updated (b7) other relevant history - updated.
|
| |
|