| Model Number |
801764 |
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/23/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Per the facility's biomedical engineer, the heart lung machine (hlm) was charged overnight but still would not run on battery.He measured the battery voltages and found them to be 12.3 volts (v) and 8.9 v.
|
| |
|
Event or Problem Description
|
|
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, the unit would not operate on battery.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported complaint was not verifiable.Multiple diligence attempts were unsuccessful for part return so an evaluation could not be completed.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
| |
|