• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1 Back to Search Results
Model Number 801764
Device Problem Failure to Run on Battery (1466)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/23/2021
Event Type  malfunction  
Manufacturer Narrative
Per the facility's biomedical engineer, the heart lung machine (hlm) was charged overnight but still would not run on battery.He measured the battery voltages and found them to be 12.3 volts (v) and 8.9 v.
 
Event Description
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, the unit would not operate on battery.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Manufacturer Narrative
The reported complaint was not verifiable.Multiple diligence attempts were unsuccessful for part return so an evaluation could not be completed.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCED PERFUSION SYSTEM 1
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
MDR Report Key11340036
MDR Text Key238621308
Report Number1828100-2021-00047
Device Sequence Number1
Product Code DTQ
UDI-Device Identifier00886799000663
UDI-Public(01)00886799000663(11)181002
Combination Product (y/n)N
PMA/PMN Number
K172220
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 05/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number801764
Device Catalogue Number801764
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/30/2021
Initial Date FDA Received02/17/2021
Supplement Dates Manufacturer Received05/03/2021
Supplement Dates FDA Received05/17/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-