Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Per the device labeling, the remedy® stemmed knee spacer may be compromised in an overweight or obese patient and/or one who does not limit the amount of activity and weight placed on the knee.The patient is obese and reported walking on the device with full weight bearing.
|
| |
|
Event or Problem Description
|
|
One month post-operative the patient was walking full weight bearing and fractured the tibial knee component.The surgeon revised the knee to remove the broken spacer and replace it with a metal tibial and femur static spacer.
|
| |
|