• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VASCULAR SOLUTIONS, LLC TRAPLINER; CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VASCULAR SOLUTIONS, LLC TRAPLINER; CATHETER Back to Search Results
Model Number 5568
Medical Device Problem Code Separation Problem (4043)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/20/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A follow-up report will be issued after the investigation is complete.
 
Event or Problem Description
As reported: while trying to advance and retrieve the trapliner through the 8f inside a guide catheter to allocate the tip into the right rca (right coronary artery), it suddenly broke and disconnected.Thetrapliner was retrieved in the guide catheter and the procedure went on without trapliner.Patient outcome is reported as well, with no problems.
 
Additional Manufacturer Narrative
Trapliner was returned for evaluation.Device history record was reviewed.There are no non-conformances related to this lot, therefore, supporting the device met material, assembly and performance specifications.The catheter pushwire was separated at the weld joint.The distal coils were herniated which is consistent with advancing/withdrawing against resistance.The ifu, states the following warning, "never advance, withdraw, or rotate an intravascular device against resistance until the cause of the resistance is determined by fluoroscopy.Movement of the catheter against resistance may result in separation of the catheter or guidewire tip, other device damage, or vessel injury." also, per ifu" exercise care in handling the catheter during a procedure to reduce the possibility of accidental breakage or kinking.Do not apply torque to the catheter during delivery, as catheter damage may result".Based on the returned product evaluation, the most likely root cause of the issue is operational context and/or unintended use error.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRAPLINER
Common Device Name
CATHETER
Manufacturer (Section D)
VASCULAR SOLUTIONS, LLC
6464 sycamore court north
minneapolis MN 55369
MDR Report Key11340661
Report Number2134812-2021-00005
Device Sequence Number16419359
Product Code DQY
UDI-Device Identifier20841156102149
UDI-Public(01)20841156102149
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K161901
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/10/2022
Device Model Number5568
Device Lot Number669470
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/01/2021
Initial Date Received by Manufacturer 01/20/2021
Supplement Date Received by Manufacturer01/20/2021
Initial Report FDA Received Date02/17/2021
Supplement Report FDA Received Date05/13/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age65 YR
Patient Weight75
-
-