• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON, S.A. BD PHASEAL OPTIMA INFUSION ADAPTER (C100-O); INTRAVASCULAR ADMINISTRATION SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON, S.A. BD PHASEAL OPTIMA INFUSION ADAPTER (C100-O); INTRAVASCULAR ADMINISTRATION SET Back to Search Results
Catalog Number 515078
Device Problem Leak/Splash (1354)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/01/2020
Event Type  malfunction  
Manufacturer Narrative
Medical device expiration date: unknown.Device manufacture date: unknown.(b)(4).Investigation summary: no photos or physical samples that display the reported condition were available for investigation.Product undergoes inspections throughout the manufacturing process to ensure the quality and functionality of the device.As the lot involved in this incident is unknown, a device history review cannot be performed and additional retained samples cannot be evaluated.Based on the available information we are not able to identify a root cause at this time.Complaints received for this device and reported condition will continue to be tracked and trended.Our quality team regularly reviews the collected data for identification of emerging trends.
 
Event Description
It was reported that the bd phaseal" optima infusion adapter (c100-o) spike broke off the bag and spilled rituxan.The following information was provided by the initial reporter: "while priming the rituxan the injector/spike came apart from the bag causing it to spill all over the medication room.Pharmacy was notified immediately".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD PHASEAL OPTIMA INFUSION ADAPTER (C100-O)
Type of Device
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
BECTON DICKINSON, S.A.
camino de valdeolivia
s/n
san agustin de guadalix
Manufacturer (Section G)
BECTON DICKINSON, S.A.
camino de valdeolivia
s/n
san agustin de guadalix
Manufacturer Contact
katie swenson
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key11340883
MDR Text Key265431466
Report Number3003152976-2021-00088
Device Sequence Number1
Product Code ONB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K181221
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other,user facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 02/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/17/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number515078
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/20/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-