Model Number TRANSDUCER X7-2
Medical Device Problem Codes Mechanical Problem (1384); Mechanical Jam (2983)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/20/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Return of the suspect transducer is anticipated.Evaluation of the transducer will be included in a follow up report upon its return and investigation completion.
 
Event or Problem Description
A customer reported an x7-2t model transducer had an articulation issue during use.There was no injury associated with this event.
 
Additional Manufacturer Narrative
A field service engineer replaced the suspect transducer for the customer to resolve the reported articulation issue.A thorough evaluation of the suspect transducer could not reproduce the articulation issue as described by the customer.The probe passed all standard testing and also passed an additional articulation test.The investigation confirmed the transducer was operating as designed with no malfunctions.
 
Event or Problem Description
A customer reported an x7-2t model transducer had an articulation issue during use.There was no injury associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X7-2T
Common Device Name
TRANSDUCER, ULTRASONIC, DIAGNOSTIC
MDR Report Key11341859
Report Number3019216-2021-10017
Device Sequence Number14289631
Product Code ITX
UDI-Device Identifier00884838075177
UDI-Public00884838075177
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K132304
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberTRANSDUCER X7-2
Device Catalogue Number989605414122
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/09/2021
Distributor Facility Aware Date01/20/2021
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer03/09/2021
Initial Report FDA Received Date02/18/2021
Supplement Report FDA Received Date04/02/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/08/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
-
-