EXACTECH, INC. EQUINOXE; MDLR CNLTD TRIDRV DRVR Back to Search Results
Catalog Number 315-25-00
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/26/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Pending evaluation.
 
Event or Problem Description
During a shoulder procedure on a male patient, the tri drive collar broke during glenoid reaming.Causing the need to change reamer handle and remove ball bearing and ring from open wound.During the same procedure, when impacting the glenoid, the glenoid impactor tip broke and also had parts removed from wound.No health complications from experience, just surgical delay.The patient was last known to be in stable condition following the event.Devices will return.Tri drive (315-25-00) stated on experience report is being captured under case (b)(4).
 
Additional Manufacturer Narrative
The disassembled tri-drive handle reported was likely the result of the retaining ring responsible for maintaining assembly of the sleeve on the modular cannulated tri-drive shaft becoming deformed as a result of obstruction of the driver collar while reaming.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EQUINOXE
Common Device Name
MDLR CNLTD TRIDRV DRVR
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 ct
gainesville FL 32653
Manufacturer Contact
kate jacobson
3523771140
MDR Report Key11342057
Report Number1038671-2021-00052
Device Sequence Number17351422
Product Code LXH
UDI-Device Identifier10885862251824
UDI-Public10885862251824
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
UNK
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number315-25-00
Device Lot Number259375025
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/29/2021
Supplement Date Received by Manufacturer10/18/2021
Initial Report FDA Received Date02/18/2021
Supplement Report FDA Received Date11/07/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexMale
-
-