• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL VISISTAT 35W 6/BOX; STAPLE, REMOVABLE (SKIN)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL VISISTAT 35W 6/BOX; STAPLE, REMOVABLE (SKIN) Back to Search Results
Model Number IPN008819
Medical Device Problem Code Mechanical Jam (2983)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/20/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The device has not been returned for investigation.Teleflex will continue to monitor and trend related events.
 
Event or Problem Description
It was reported that (b)(6) 2021, the staple was found jamming and could not be fired prior to the patient.
 
Event or Problem Description
It was reported that on (b)(6) 2021, the staple was found jamming and could not be fired prior to the patient.
 
Additional Manufacturer Narrative
(b)(4).Per dhr the product visistat 35w 6/box lot#: 73c1900104 was manufactured on 03/05/2019 a total of (b)(4) pieces.Lot was released on 03/13/2019.Dhr investigation did not show issues related to complaint.The customer returned one unit 528235 visistat 35w 6/box for investigation.The returned stapler was visually examined with and with out magnification.Visual examination of the returned stapler revealed that the sample had its trigger partially engaged.The staples appeared to be properly aligned.The stapler was received with 37 staples left in the device.Functional inspection was performed by attempting to fire staples from the device.Using hand pressure, the trigger was engaged in order to complete the trigger cycle.Upon full engagement of the trigger, the first staple was able to properly form and release from the device.These actions were repeated for the next 4 staples with the same result.To simulate insertion into the skin, the stapler was fired into a simulated skin pad.All 32 remaining staples were able to properly fire and close into the skin pad.All staples were fired from the device with no difficulty.No defects or anomalies were observed.The ifu for this product, l02644, was reviewed as a part of this complaint investigation.The ifu states "to obtain optimum staple c losure, the trigger must be squeezed all the way in." a corrective action is not required at this time as there were no functional issues found with the returned device.The stapler was able to properly fire all remaining staples.The reported complaint of "misfire/jamming - staples not firing" could not be confirmed based upon the sample received.The returned stapler was able to form and release all remaining staples in the cover block.No defects or anomalies were observed with the sample.A device history record review was performed on the device with no evidence to suggest a manufacturing related cause.There were no functional issues found with the returned stapler.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISISTAT 35W 6/BOX
Common Device Name
STAPLE, REMOVABLE (SKIN)
Manufacturer (Section D)
TELEFLEX MEDICAL
morrisville NC
MDR Report Key11342243
Report Number3003898360-2021-00189
Device Sequence Number12011218
Product Code GDT
UDI-Device Identifier04026704600762
UDI-Public04026704600762
Combination Product (Y/N)N
Initial Reporter CountryCH
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberIPN008819
Device Catalogue Number528235
Device Lot Number73C1900104
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/17/2021
Initial Date Received by Manufacturer 01/21/2021
Supplement Date Received by Manufacturer03/09/2021
Initial Report FDA Received Date02/18/2021
Supplement Report FDA Received Date03/09/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
N/A.
-
-