BERLIN HEART GMBH BLOOD PUMP PU VALVES 15 ML, IN/OUT 9 MM; VENTRICULAR ASSIST DEVICE Back to Search Results
Model Number P15P-001
Device Problem Output Problem (3005)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/29/2021
Event Type  malfunction  
Manufacturer Narrative
The excor blood pump, s/n (b)(4) was in use by the patient from 2020-03-12 until 2021-01-29 (323 days).We have reviewed the production records of the excor blood pump, s/n (b)(4).This pump was produced according to our specification.The blood pump is designed with a triple layer membrane separating the air chamber from blood chamber for safety reasons.The entire membrane consists of an air-side layer, a middle layer and a blood-side layer.In case of disruption in one of the triple layers, there are two more layers that will maintain the integrity of the air and blood chambers.A detailed investigation report will be provided as soon as it is available.
 
Event Description
Berlin heart was contacted by the site to report a suspected membrane defect in the excor blood pump of a patient supported in the lvad configuration due to an incomplete ejection of the affected blood pump.Berlin heart recommended a pump exchange.The affected blood pump was exchanged in the clinic by trained personnel.The exchange was performed without complications and the patient is doing well.
 
Manufacturer Narrative
During initial visual examination of the returned blood pump, an air cushion could be detected between the membrane layers.The blood pump was then tested for functional performance where it did not reach its required functional specification, confirming the customer's complaint.For further analysis, the pump was disassembled, and the membrane layers were individually examined.A leakage was detected in the air-side layer along the rolling radius of the stabilization ring.Furthermore, graphite abrasion was found between the membranes.The middle layer and the blood-side layer of the triple layer membrane were found to be intact.The thickness of the defective membrane layer, and the one adjacent was re-measured at defined points.The thickness of the individual layers at all defined points and in the area around the leakage was found to be within specification at the time of the re-measurement.The cause of the leakage in the air-side layer was most likely a graphite abrasion between the membrane layers.This caused increased friction at some points, which finally led to the defect in the air-side layer of the triple-layer membrane.As a result of this defect, air got in between the membrane layers and formed an air cushion, causing the reduced pump performance (incomplete filling and emptying).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BLOOD PUMP PU VALVES 15 ML, IN/OUT 9 MM
Type of Device
VENTRICULAR ASSIST DEVICE
Manufacturer (Section D)
BERLIN HEART GMBH
wiesenweg 10
berlin, 12247
GM  12247
MDR Report Key11342673
MDR Text Key232362752
Report Number3004582654-2021-00011
Device Sequence Number1
Product Code DSQ
UDI-Device Identifier04260090040157
UDI-Public04260090040157
Combination Product (y/n)N
PMA/PMN Number
P160035
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Remedial Action Replace
Type of Report Initial,Followup
Report Date 02/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/18/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2022
Device Model NumberP15P-001
Device Catalogue NumberP15P-001
Was Device Available for Evaluation? No
Date Manufacturer Received01/29/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age2 YR
-
-