Model Number ARD568603999
Device Problems Break (1069); Detachment of Device or Device Component (2907)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Additional information will be provided following the conclusion of the investigation.
 
Event Description
On (b)(6) 2021 getinge became aware of an issue with lucea 100 surgical light.As it was stated, the housing was broken.Photographic evidence shows that plastic particles were missing.There was no injury reported, however, we decided to report the complaint in abundance of caution, as any particles falling off into sterile field or during procedure may cause contamination.
 
Event Description
Manufacturer reference number (b)(4).
 
Manufacturer Narrative
On 15th february, 2021 getinge became aware of an issue with lucea 100 surgical light.As it was stated, the housing was broken.Photographic evidence shows that plastic particles were missing.There was no injury reported, however, we decided to report the complaint in abundance of caution, as any particles falling off into sterile field or during procedure may cause contamination.It was established that when the event occurred, the surgical light did not meet its specification as it was damaged and particles were missing, and it contributed to the event.None of the provided information indicate that upon the event occurrence the device was being used for patient treatment.During the investigation it was found that in the past the reported scenario has never lead to serious injury or worse, to death.Cover cracking is indicated by our product experts to likely be caused by combination of different factors such as: mechanical stress, environmental conditions (such a high temperature and humidity variations during transport and storage), use of inappropriate cleaning/disinfections products, or inappropriate cleaning and disinfection protocols.We believe that all remaining devices are performing correctly in the market.We also believe that if the preventive maintenance would have been followed the incident could have been avoided.Given the circumstances we shall continue to monitor for any further events of this nature and do not propose any further action at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LUCEA 100
Type of Device
LIGHT, SURGICAL, CEILING MOUNTED
Manufacturer (Section D)
MAQUET SAS
parc de limere
avenue de la pomme de pin
ardon
MDR Report Key11343513
MDR Text Key232393366
Report Number9710055-2021-00063
Device Sequence Number1
Product Code FSY
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 03/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/18/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberARD568603999
Device Catalogue NumberARD568603999
Was Device Available for Evaluation? Yes
Date Manufacturer Received02/18/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-