Model Number 1550300-23
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Codes Ischemia (1942); Myocardial Infarction (1969)
Date of Event 12/17/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The stent remains in patient.The device will not be returned for evaluation.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
On (b)(6) 2019, the patient had a 3.0x23mm xience sierra implanted in the left anterior descending artery with no issues.On (b)(6) 2020 the patient experienced a heart attack due to blockage from the xience stent becoming separated into different pieces.Some blood flow was coming through, but not a lot.A quadruple bypass was performed on the (b)(6) to treat the heart attack.The separated stent still remains in the patient after the surgery.There was no adverse patient sequela and no clinically significant delay.No additional information was provided.
 
Additional Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar incidents from this lot.The investigation was unable to determine a conclusive cause for the reported material separation (stent).A conclusive cause for the reported patient effects, and the relationship to the product, if any, cannot be determined; however, the subsequent treatment appears to be related to circumstances of the procedure.The reported patient effects of ischemia and myocardial infarction are listed in the xience sierra, everolimus eluting coronary stent system (eecss), electronic instructions for use (eifu) as known patient effects of coronary stenting procedures there is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE SIERRA EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Common Device Name
DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key11343518
Report Number2024168-2021-01290
Device Sequence Number14403306
Product Code NIQ
UDI-Device Identifier08717648227271
UDI-Public08717648227271
Combination Product (Y/N)Y
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
P110019
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,consum
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 03/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/23/2020
Device Model Number1550300-23
Device Catalogue Number1550300-23
Device Lot Number9061741
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 01/26/2021
Supplement Date Received by Manufacturer03/03/2021
Initial Report FDA Received Date02/18/2021
Supplement Report FDA Received Date03/25/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention; Disability;
Patient Age35 YR
Patient Weight88
-
-