• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GYRUS ACMI, INC BURR, 3MM, 70DEG, DIAMOND BULLET, 5/PK; EAR, NOSE, AND THROAT BUR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GYRUS ACMI, INC BURR, 3MM, 70DEG, DIAMOND BULLET, 5/PK; EAR, NOSE, AND THROAT BUR Back to Search Results
Model Number BUR3070DB
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/01/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The subject device referenced in this report was not returned for evaluation, therefore the device evaluation could not be completed at this time.The exact cause of the reported event could not be conclusively determined at this time.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
 
Event or Problem Description
The user facility reported that during a functional endoscopic sinus surgery (fess), the device did not work as intended.The burr had initially worked, then eventually stopped completely.A console and two (2) separate handpieces were used with the burr for testing, however the issue remained.The drill was taken from the handpiece and it was observed that part of the white piece of plastic in the drill where the drill is inserted into the handpiece was chipped up.The procedure was prolonged over an hour and was then aborted due to a lack of burrs.The patient had to be rescheduled for a new procedure.There was no patient harm reported.The device was inspected prior to the procedure and no damage or abnormalities were observed.The user facility later reported that the reason why the burrs broke was that they were used on a patient with an osteoma.The bone tissue was too hard for the burr to drill away.The rotation speed of the shaver became too small to handle this.The customer acknowledged that burrs were used in a procedure they are not intended to be used in.This complaint is related to patient identifiers; (b)(6) and (b)(6).
 
Additional Manufacturer Narrative
This supplemental report is being submitted to provide additional information that was received from the user facility and the legal manufacturer¿s investigation.Per the customer, there were no comorbidities as the patient was a healthy young man.The cause of the surgery was due to a large osteoma that grew into the orbit and resulted in reduced mobility and double vision.Everything went as planned during the surgery except that the equipment did not work and they did not get the last part of the osteoma.The legal manufacturer performed an investigation.The dhr review could not be performed as no lot number was reported and there was no device return to inspect for a lot number.If the lot number becomes available at a later date, the dhr shall be reviewed.The definitive root cause of the burr tightness could not be confirmed as the device was not returned.However, per the torque verification procedure for diego elite burrs, a burr undergoes a test where the burr is smoothly rotated until a click is heard, or induction weld fails.As a result, it is unlikely that the burr would be shipped with issues regarding the burr feeling tight or not spinning.This suggests that the damages that may have been incurred on the burr were a result of transportation or use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BURR, 3MM, 70DEG, DIAMOND BULLET, 5/PK
Common Device Name
EAR, NOSE, AND THROAT BUR
Manufacturer (Section D)
GYRUS ACMI, INC
136 turnpike road
southborough MA 01772
MDR Report Key11344179
Report Number1037007-2021-00017
Device Sequence Number14711289
Product Code EQJ
UDI-Device Identifier00821925038257
UDI-Public00821925038257
Combination Product (Y/N)N
Initial Reporter CountryNO
PMA/510(K) Number
EXEMPT-EQJ
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 03/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberBUR3070DB
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 02/02/2021
Supplement Date Received by Manufacturer02/22/2021
Initial Report FDA Received Date02/18/2021
Supplement Report FDA Received Date03/12/2021
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MDCONS100, MULTIDEBRIDER POWER CONSOLE.; MDHP100S, MULTIDEBRIDER HANDPIECE, STRAIGHT.; MDHP100S, MULTIDEBRIDER HANDPIECE, STRAIGHT.
-
-