• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHROCARE CORP. FIRSTPASS SUTURE PASSER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHROCARE CORP. FIRSTPASS SUTURE PASSER Back to Search Results
Model Number 22-4035
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/25/2021
Event Type  Injury  
Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event Description
It was reported that during the surgery when the suture was being retrieved, after the bite the firstpass suture passer upperjaw broke off.No significant delay or other complications reported.Smith and nephew back-up device was available to complete the surgery.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
H10 h3,h6: the reported device, used in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was an isolated issue.A review of the instructions for use found that excessive force can result in device damage.A review of risk management files found that the reported failure was documented appropriately.A relationship, if any, between the subject device and the reported event could not be determined.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.There were no indications that would suggest that the device did not meet product specifications upon release into distribution.If the product associated with this event is returned at a future date, this evaluation will be reopened for investigation.
 
Manufacturer Narrative
H10 h3,h6: the reported device, used in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection of the returned instrument shows a detached suture capture.No manufacturing abnormalities.A functional evaluation revealed the needle will deploy when trigger is initiated, the suture capture is detached and has been returned.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review concluded this was an isolated event.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A review of risk management files found that the reported failure was documented appropriately.The complaint was confirmed.Factors that could have contributed to the reported event include: (1) excessive force (2) tissue thickness (3) damage or debris on the device tip between passes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FIRSTPASS SUTURE PASSER
Type of Device
PASSER
Manufacturer (Section D)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
MDR Report Key11345150
MDR Text Key232507385
Report Number3006524618-2021-00219
Device Sequence Number1
Product Code HWQ
UDI-Device Identifier00817470000160
UDI-Public00817470000160
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,other
Type of Report Initial,Followup,Followup
Report Date 05/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/18/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number22-4035
Device Catalogue Number22-4035
Device Lot NumberK5001CR
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/24/2021
Date Manufacturer Received05/10/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-