|
It was reported that during the resection procedure of the larynx lesion using the procise lw wand, an lw radiofrequency tip was used, after 5 min of use the electrode broke off and the tip stopped working, did not coagulate.The procedure was successfully completed with a delay of 30 minutes or less using a back-up device.No patient injury or other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
|
H10 h3,h6: the reported device, used in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was an isolated event.A review of the instructions for use found the following statement: "the wand is supplied sterile, and is for single use only.Do not clean, resterilize, or reuse the wand, as this may damage or compromise the performance of the device resulting in product malfunction, failure, or patient injury." a review of risk management files found that the reported failure was documented appropriately.A relationship, if any, between the subject device and the reported event could not be determined.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.If the product associated with this event is returned at a future date, this evaluation will be reopened for investigation.
|