SPACELABS HEALTHCARE (WASHINGTON), INC TRULINK; SYSTEM, MEASUREMENT, BLOOD-PRESSURE, NON-INVASIVE
|
Back to Search Results |
|
Model Number 90496 |
Device Problem
Therapeutic or Diagnostic Output Failure (3023)
|
Patient Problem
Insufficient Information (4580)
|
Event Date 02/05/2021 |
Event Type
malfunction
|
Event Description
|
Monitor unable to take blood pressure (bp) despite changing b/p cable, cuff and monitor box.No patient harm.(this was one of a set of twins) see other twin report same day.Biomed in-house findings: found inflation error when trying to test nibp (non-invasive blood pressure) because of leaky bp hose.Replaced with known working bp hose.Successfully tested nibp.
|
|