| Model Number |
269-100-070 |
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
01/26/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The returned device is currently under investigation.A follow up report with results of the investigation will be submitted upon completion.
|
| |
|
Event or Problem Description
|
|
After removing the device from patient, the surgeon noticed metal debris and damage to the instrument.
|
| |
|
Additional Manufacturer Narrative
|
|
An evaluation of the 70mm blade determined it had experienced a failure of the right jaw component where it assembles with the hinge-pin.This failure mode is consistent with excessive loading of the cross-sectional area between the pin and the jaw.This excessive loading is brought on by over-tightening the blade during initial assembly of the blade to the screw shank component prior to insertion into the surgical site.The stress fracture is then propagated to a full break when the blade is subjected to normal forces during the procedure.All broken pieces of the blade were retrieved from the surgical site, and the or team has been notified about the potential complications of over-tightening during assembly.
|
| |
|