• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US REUSE PATIENT SET FMS 24PK; SURGICAL IRRIGATION TUBING SET, GENERAL-PURPOSE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US REUSE PATIENT SET FMS 24PK; SURGICAL IRRIGATION TUBING SET, GENERAL-PURPOSE Back to Search Results
Model Number 284649
Medical Device Problem Codes Crack (1135); Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/05/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4) incomplete the lot number was unknown.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Event or Problem Description
It was reported by the customer in (b)(6) that during an arthroscopic shoulder stabilization procedure on 2/5/2021, it was observed that the irrigation pump device had a water leak in the hose that it was split in the center.There were no adverse patient consequences reported.No additional information was provided.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: investigation summary : according to the information provided, it was reported that the tube is connected to the irrigation pump code 284649, it works, the pressure is increased to purge and it fails, it is disconnected and reconnected and the valve is filled.Again it fails and then is disconnected, the tube was checked and there was a water leak in the hose, when checking it, it was observed that it was split in the center, so it could no longer be used.The complaint device is not being returned, therefore unavailable for a physical evaluation.However 2 photos were provided.Upon reviewing the photos, it could be observed that a section of the inflow tubing is damaged, a slit could be observed near the expansion chamber.As a potential cause cannot be associated to manufacturing, therefore a manufacturing record evaluation is not required.As part of depuy synthes mitek quality process all devices are manufactured, inspected, and released to approved specifications.According with the visual inspection result, this complaint can be confirmed.A possible root cause can be attributed to the handling of the device at the moment when the tubing is placed into the roller pump, the tubing could have grabbed by the edge of the rolling part.As per ifu 111204: to avoid damage to tubing, make sure that the tubing is centered on the groove of the pump.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to monitor additional complaint information for potential safety signals through complaint trending as part of post market surveillance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REUSE PATIENT SET FMS 24PK
Common Device Name
SURGICAL IRRIGATION TUBING SET, GENERAL-PURPOSE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key11349316
Report Number1221934-2021-00603
Device Sequence Number3416736
Product Code HRX
UDI-Device Identifier10886705023233
UDI-Public10886705023233
Combination Product (Y/N)N
Initial Reporter CountryMX
PMA/510(K) Number
K951843
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number284649
Device Catalogue Number284649
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/05/2021
Supplement Date Received by Manufacturer04/11/2021
Initial Report FDA Received Date02/19/2021
Supplement Report FDA Received Date04/12/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-