BOSTON SCIENTIFIC CORPORATION STERLING; CATHETER, PERCUTANEOUS Back to Search Results
Model Number 24716
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/29/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device evaluated by mfr.: the device was returned for analysis.The outer shaft, inner shaft, balloon and tip were visually and microscopically examined.Visual examination revealed no damages.Microscopic examination revealed a pinhole in the guidewire lumen 21cm from the tip.Inspection of the remainder of the device presented no other damage or irregularities.
 
Event or Problem Description
Reportable based on device analysis completed on 11-feb-2021.It was reported that balloon damage occurred.The chronic totally occluded target lesion was located in the moderately calcified superficial femoral artery.A 6.0x220x150-hybrid sterling balloon catheter was advanced for dilation.However, on initial inflation under fluoroscopy, the balloon was twisting approximately 5cm from the proximal marker.The device was removed and the procedure was completed with another of the same device.No patient complications nor harm were reported.However, returned device analysis revealed shaft hole.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STERLING
Common Device Name
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key11349787
Report Number2134265-2021-01920
Device Sequence Number4117123
Product Code LIT
UDI-Device Identifier08714729845485
UDI-Public08714729845485
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
K132430
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/08/2021
Device Model Number24716
Device Catalogue Number24716
Device Lot Number0025505981
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/14/2021
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/11/2021
Initial Report FDA Received Date02/19/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-