• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHROCARE CORP. MICROBLATOR 30 ICW; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHROCARE CORP. MICROBLATOR 30 ICW; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number AC4050-01
Medical Device Problem Code Failure to Deliver Energy (1211)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/02/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Reference: (b)(4).
 
Event or Problem Description
It was reported that, during surgery, the microblator failed to coagulate and the tip did not work at that time.The procedure was completed without significant delay using a competitor device.No patient injury or other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
H10 h3,h6: the reported device, used in treatment, was received for evaluation.There was no relationship found between the returned device and the reported incident.A visual inspection of the returned instrument shows no manufacturing abnormalities.No visual issues were observed.The device was plugged into the controller and registered settings (1,6).The wand was able to generate plasma as intended.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review concluded this was an isolated event.A review of the instructions for use found: -use only conductive media (e.G.Saline, ringer¿s lactate, etc.).Do not use non-conductive media (e.G.Sterile water, air, gas, glycine, etc.).-under arthroscopic guidance, ensure that the wand tip is completely surrounded by irrigant solution during use.-when using a wand, ensure adequate flow and circulation of saline solution to prevent unnecessary heating of the solution that may result in tissue damage.-do not contact metal objects while activating the wand.Damage to the tip may result -microblator 30 wand with integrated cable is designed for ablation and/ or coagulation only, and not for mechanical displacement of tissue through applied force which may result in bent or detached electrodes.A review of risk management files found that the reported failure was documented appropriately.The complaint was not verified and the root cause could not be determined since the reported malfunction could not be duplicated during the product evaluation process.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICROBLATOR 30 ICW
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
MDR Report Key11351178
Report Number3006524618-2021-00224
Device Sequence Number4045178
Product Code GEI
UDI-Device Identifier00817470003123
UDI-Public00817470003123
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K033584
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/01/2023
Device Model NumberAC4050-01
Device Catalogue NumberAC4050-01
Device Lot Number2053039
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/23/2021
Initial Date Received by Manufacturer 02/02/2021
Supplement Date Received by Manufacturer03/26/2021
Initial Report FDA Received Date02/19/2021
Supplement Report FDA Received Date03/27/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-