• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LEMAITRE VASCULAR, INC. LEMAITRE AORTIC OCCLUSION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LEMAITRE VASCULAR, INC. LEMAITRE AORTIC OCCLUSION CATHETER Back to Search Results
Model Number 2107-81
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 01/08/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
We have not received the complaint device for investigation.Without the returned device, we remain inconclusive about the root cause of this malfunction.The device was used in a cadaver during an organ explant.Surgeon was able to explant the organ in a viable condition.However, surgeon needed to speed up the procedure since the entire procedure must be completed in less than 25 minutes.Our review of the lot history records for this lot did not find any discrepancies either in the manufacturing or packaging process that could be related to this incident.Please note that we have also submitted manufacturer incident report# 1220948-2021-00009 related to another lemaitre aortic occlusion incident that occurred during the same procedure.
 
Event or Problem Description
Surgeon used lemaitre aortic occlusion catheter (ref# 2107-81 lot# olc1059) on a deceased patient during organ explant.However, the balloon of this catheter deflated during the procedure.Despite the malfunction, surgeon was able to remove the organ successfully and just in time and as he was able to complete the entire procedure in less than 25 minutes.This is report 2 of 2.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEMAITRE AORTIC OCCLUSION CATHETER
Common Device Name
OCCLUSION CATHETER
Manufacturer (Section D)
LEMAITRE VASCULAR, INC.
63 second ave
burlington MA 01803
MDR Report Key11352882
Report Number1220948-2021-00010
Device Sequence Number3545412
Product Code MJN
UDI-Device Identifier00840663101658
UDI-Public00840663101658
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K132022
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Pharmacist
Type of Report Initial
Report Date (Section B) 02/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/28/2024
Device Model Number2107-81
Device Catalogue Number2107-81
Device Lot NumberOLC1059
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/20/2021
Initial Report FDA Received Date02/19/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-