| Medical Device Problem Codes |
Vibration (1674); Noise, Audible (3273)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.The actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the reported condition that the attachment device had ¿shaking method¿ and rattle was confirmed.An assessment was performed and it was determined that the device had vibration.It was further determined that the device failed pretest for vibration assessment.The assignable root cause of this condition was determined to be traced to maintenance, which is user error/misuse/abuse.Udi :(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during pre-surgery testing it was observed that the attachment device had ¿shaking method¿ and rattle.During in-house engineering evaluation it was determined that the device had vibration and was unable to be disassembled.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention, or prolonged hospitalization.The exact date of the event was not reported.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
| |
|