SMITH & NEPHEW, INC. QUICK CLAMP 10.5MM TO 5MM; APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBO, MULTIPLE COMPONENT
|
Back to Search Results |
|
| Model Number |
71067372 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/31/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the surgeon was tightening the nut on the clamp, outside the patient, then the clamp broke and fell into pieces.No delay or injury reported.The procedure was finished using a smith and nephew backup device.
|
| |
|
Additional Manufacturer Narrative
|
|
Results of investigation: the device, intended for use in treatment, was not returned for evaluation.Therefore, product analysis could not be performed at this time.So the reported event could not be confirmed.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|
Search Alerts/Recalls
|
|
|