| Model Number |
8115539 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/21/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Country: (b)(4).Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Information was received a healthcare provider via manufacturer representative regarding a patient with an indication of degenerative disease in need of lumbar revision surgery used in spinal therapy.It was reported that when implanted the crosslink in the patient's body, the screw plug broke, and the doctor immediately took out the broken crosslink without causing any damage to the patient.The patient has recovered and discharged. there was no delay reported.There were no patient symptoms reported.There were no fragments in the patient reported.There were no further complications reported regarding the event.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation summary: visual and optical inspection did not reveal any cracks or deformation to the crosslink.No signs of inner or outer thread damage.The hex portion of the one of the set screws has been sheared off.Functional inspection of the center nut confirmed the nut was able to be tightened and loosened without any issue.This type of damage is consistent with torsional overload.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|