| Model Number |
7205682S |
| Medical Device Problem Code |
Crack (1135)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/19/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference case (b)(4).The reported device, used in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection found distal tip and fiber damage and cracked distal lens with scratches.The complaint was confirmed.Factors that could have contributed to the reported event include an impact event inconsistent with normal use.
|
| |
|
Event or Problem Description
|
|
It was reported that during knee arthroscopy procedure, the scope had the lens cracked.No delay and a back up was available to complete the procedure.No other complications were reported.Results of investigation have concluded that this scope had a cracked on the distal lenses which makes it a reportable event.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
| |
|
Search Alerts/Recalls
|