• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. PENUMBRA SMART COIL; HCG, KRD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. PENUMBRA SMART COIL; HCG, KRD Back to Search Results
Model Number 400SMTSFT0615
Medical Device Problem Code Failure to Advance (2524)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/22/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.This report is associated with mfr report numbers: 3005168196-2021-00314, 3005168196-2021-00316,.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure to treat a neuro fistula using penumbra smart coils (smart coils) and a non-penumbra microcatheter.During the procedure, three smart coils would not advance within their introducer sheaths; therefore, the three smart coils were removed.The procedure was completed using additional smart coils and the same microcatheter.There was no report of an adverse effect to the patient.
 
Additional Manufacturer Narrative
Evaluation of the returned smart coil revealed that the pusher assembly mid-joint was retracted proximal to the introducer sheath friction lock.If this occurs, resistance will likely be experienced during advancement and the device may not be advanced out of the introducer sheath.During functional testing, the pusher assembly mid-joint was advanced through the introducer sheath friction lock with resistance.Then the smart coil was able to advance through the introducer sheath without issue.Further evaluation of the device revealed a pusher assembly kink.This damage was likely incidental to the complaint and may have occurred during packaging for return to penumbra.Evaluation of the returned smart coil revealed that the pusher assembly was kinked, and the embolization coil had offset coil winds.If the device is forcefully advanced against resistance, damage such as this may occur.The root cause of the resistance could not be determined.During functional testing, the smart coil was able to advance through the introducer sheath without issue.Evaluation of the returned smart coil revealed that the pusher assembly mid-joint was retracted proximal to the introducer sheath friction lock.If this occurs, resistance will likely be experienced during advancement and the device may not be advanced out of the introducer sheath.Forceful advancement against this resistance may have contributed to the kinks and fracture which were observed near the pusher assembly mid-joint.Further evaluation of the device revealed an additional pusher assembly kink.This damage was likely incidental to the complaint and may have occurred during packaging for return to penumbra.Penumbra coils are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.Section h.Box 6.Conclusions code 1: 4316 - the investigation findings do not lead to a clear conclusion about the root cause of the resistance during the procedure.This report is associated with mfr report numbers: 1.3005168196-2021-00314.2.3005168196-2021-00316.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SMART COIL
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key11359925
Report Number3005168196-2021-00315
Device Sequence Number16980102
Product Code HCG
UDI-Device Identifier00814548014975
UDI-Public00814548014975
Combination Product (Y/N)Y
Initial Reporter CountryCA
PMA/510(K) Number
K160832
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/25/2021
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/24/2025
Device Model Number400SMTSFT0615
Device Catalogue Number400SMTSFT0615
Device Lot NumberF99769
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/29/2021
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 01/25/2021
Supplement Date Received by Manufacturer05/20/2021
Initial Report FDA Received Date02/22/2021
Supplement Report FDA Received Date05/26/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Date Report Sent to FDA01/01/2005
-
-