• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN LP LLC NORTH HAVEN ENDO GIA; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN LP LLC NORTH HAVEN ENDO GIA; STAPLE, IMPLANTABLE Back to Search Results
Model Number 030452
Medical Device Problem Codes Break (1069); Entrapment of Device (1212)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/25/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during thoracoscopic lobectomy, at the time of bronchial resection, the part of the reload that connects to the handle broke.The jaws were locked on the tissue.The user pulled the black return knob, opened the jaws, and removed the jaws normally.Another handle and reload were used to resolve the issue in order to complete the case.There was no patient injury.
 
Additional Manufacturer Narrative
H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the reload had all full complement of staples and the jaw of the reload was open.The adapter of the reload was broken and one part came off.It was reported that the device broke.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.This issue may occur due to rough handling or excessive manipulation.It was also reported that the jaws of the device remained closed on tissue after firing.The reported issue could not be confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: preoperative radiotherapy may result in changes to tissue.These changes may, for example, cause the tissue thickness to exceed the indicated range for the selected staple size.Careful consideration should be given to any pre-surgical treatment the patient may have undergone and in corresponding selection of staple size.Appropriate tissue thickness ranges for cartridge sizes.When positioning the stapler on the application site, ensure that no obstructions, such as clips, are incorporated into the instrument jaws.Firing over an obstruction may result in incomplete cutting action and/or improperly formed staples.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDO GIA
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
MDR Report Key11361200
Report Number1219930-2021-00707
Device Sequence Number10960089
Product Code GDW
UDI-Device Identifier10884523002836
UDI-Public10884523002836
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K111825
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number030452
Device Catalogue Number030452
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/01/2021
Initial Date Received by Manufacturer 01/27/2021
Supplement Date Received by Manufacturer03/02/2021
Initial Report FDA Received Date02/22/2021
Supplement Report FDA Received Date03/24/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-