• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION WATCHMAN LAA CLOSURE DEVICE & DELIVERY SYSTEM; SYSTEM, APPENDAGE CLOSURE, LEFT ATRIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION WATCHMAN LAA CLOSURE DEVICE & DELIVERY SYSTEM; SYSTEM, APPENDAGE CLOSURE, LEFT ATRIAL Back to Search Results
Model Number 10371
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Pericardial Effusion (3271)
Date of Event 02/04/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported a pericardial effusion occurred.A left atrial appendage (laa) closure procedure was performed.A 30mm watchman laa closure device was successfully implanted.The patient was on xarelto.The following morning at 8 a.M., a small pericardial effusion was noticed.There were no signs of tamponade.The patient's blood pressure changed and was treated with midodrine.Echocardiograms were performed daily and a transthoracic echocardiogram showed a possible mild increase in the size of the effusion on the subcostal views.The patient was discharged home four days post procedure and will be followed as outpatient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WATCHMAN LAA CLOSURE DEVICE & DELIVERY SYSTEM
Common Device Name
SYSTEM, APPENDAGE CLOSURE, LEFT ATRIAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key11362602
Report Number2134265-2021-02160
Device Sequence Number19094241
Product Code NGV
UDI-Device Identifier08714729838258
UDI-Public08714729838258
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P130013
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/16/2022
Device Model Number10371
Device Catalogue Number10371
Device Lot Number0024121654
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/08/2021
Initial Report FDA Received Date02/22/2021
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-