• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BULB TYPE IRRIGATION SYRINGE 50ML, PLASTIC STERILE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BULB TYPE IRRIGATION SYRINGE 50ML, PLASTIC STERILE Back to Search Results
Model Number 0035280
Medical Device Problem Code Device Contamination with Chemical or Other Material (2944)
Health Effect - Clinical Codes No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/25/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event or Problem Description
It was reported that a foreign material was found in a sterilized pack before opening the package.
 
Additional Manufacturer Narrative
The reported event was confirmed however the cause was unknown.One sample was confirmed to exhibit the reported failure.The device had not met specifications.Visual evaluation of the returned sample noted one opened (without original packaging), unused bulb syringe.Visual inspection of the sample noted that there was a 2.50 sq mm green silicone-like material within packaging, but outside of the syringe.This does not meet the specification as "product and package (kit) must be free from visible loose or embedded foreign matter greater than an aggregate total of 0.6mm2 or 1/16¿ in length.A potential root cause for this failure mode could be ¿no following to production areas cleaning procedure¿.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.Labeling review was not performed because labeling could not have prevented the reported failure.Corrections: d, h.H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the actual/suspected device was inspected.
 
Event or Problem Description
It was reported that a foreign material was found in a sterilized pack before opening the package.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BULB TYPE IRRIGATION SYRINGE 50ML, PLASTIC STERILE
Common Device Name
BULB TYPE IRRIGATION SYRINGE
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
MDR Report Key11362988
Report Number1018233-2021-00679
Device Sequence Number823532
Product Code KYZ
UDI-Device Identifier00801741049255
UDI-Public(01)00801741049255
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other,user facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2025
Device Model Number0035280
Device Catalogue Number0035280
Device Lot NumberNGEN4086
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/18/2021
Initial Date Received by Manufacturer 01/27/2021
Supplement Date Received by Manufacturer07/23/2021
Initial Report FDA Received Date02/22/2021
Supplement Report FDA Received Date08/16/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-