Model Number 55890006550
Medical Device Problem Code Mechanics Altered (2984)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/02/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Information was received from the healthcare professional (hcp) via the manufacturer representative regarding an event which occurred during posterior slip reset procedure in a patient diagnosed with spondylolisthesis.It was reported that the spacer broke, the sc rewdriver's holding pattern was damaged and it was impossible to screw and the set screws slipped.Cage was pulled out with tools on the spot, no secondary surgery.Products will be returned.No patient injury / complication was reported.
 
Additional Manufacturer Narrative
H3: product analysis: part # 55890006550, lot # h5578421 visual and optical inspection revealed some damage to the female torx.The crown of the screw has been worn from the seating of the rod.Functional check with a sample set screw confirmed the set screw was able to thread into the head of the screw and the head was still able to pivot without any issue.The screw appears to function as intended.The female torx appears to be damaged from the removal process.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CD HORIZON SPINAL SYSTEM
Common Device Name
APPLIANCE, FIXATION, SPINAL INTERLAMINAL
Manufacturer (Section D)
MDT SOFAMOR DANEK PUERTO RICO MFG
barrio marianna rd 909, km0.4
humacao PR 00792
MDR Report Key11363848
Report Number1030489-2021-00243
Device Sequence Number13686655
Product Code KWP
UDI-Device Identifier00643169058170
UDI-Public00643169058170
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K113174
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/06/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number55890006550
Device Catalogue Number55890006550
Device Lot NumberH5578421
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/18/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/03/2021
Supplement Date Received by Manufacturer07/16/2021
Initial Report FDA Received Date02/23/2021
Supplement Report FDA Received Date08/06/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age34 YR
-
-