• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: REVERSE MEDICAL MICROVASCULAR PLUG; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

REVERSE MEDICAL MICROVASCULAR PLUG; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION Back to Search Results
Model Number MVP-9Q
Medical Device Problem Codes Migration or Expulsion of Device (1395); Activation, Positioning or Separation Problem (2906)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/01/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Physician was attempting to use microvascular plug during procedure to treat a severely calcified lesion in the proximal common iliac artery.The device was inspected with issues noted.Device would not expand anymore.The device was prepped per ifu with issues identified.It was reported that the mvp was intended to use to embolize a hypo gastric artery.The mvp was put in the hypo gastric artery through a non-medtronic 5fr 65cm sheath and 0.038" angiographic catheter (2 side holes).After placement, the catheter including the mvp dislocated due to tension in the system (retrograde approach).The mvp was retrieved and then found out that the mvp would not expand anymore to its actual size.Physician replaced it with a new product and successfully completed the procedure.There was no patient injury.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICROVASCULAR PLUG
Common Device Name
DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
Manufacturer (Section D)
REVERSE MEDICAL
13700 alton parkway
ste. 167
irvine CA 92618
MDR Report Key11363915
Report Number3007170829-2021-00004
Device Sequence Number11623860
Product Code KRD
UDI-Device Identifier00847536027732
UDI-Public00847536027732
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
K150108
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/28/2021
Device Model NumberMVP-9Q
Device Catalogue NumberMVP-9Q
Device Lot NumberA920382
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/16/2021
Supplement Date Received by Manufacturer09/20/2021
Initial Report FDA Received Date02/23/2021
Supplement Report FDA Received Date09/20/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-