| Model Number |
110-120 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
10/27/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side rupture.The device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: visual analysis of the returned device identified: underweight, red encapsulation, fold creases, and an extended opening on posterior.A microscopic analysis was performed which identified: a sharp opening on posterior, two striated openings on anterior, and wear abrasion.A dimension measurement in the shell was performed which identify the thickness within specification.Based on the device analysis the final assessment is: - a sharp opening on posterior assessed as unidentified (tear) opening.- two striated openings on anterior assessed as surgical damage consist in the use of some surgical tool.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side rupture.The device has been explanted.
|
| |
|