| Catalog Number |
301-05-01 |
| Medical Device Problem Codes |
Break (1069); Fracture (1260)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/03/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Pending evaluation.
|
| |
|
Event or Problem Description
|
|
As reported, during the impaction of a preserve stem broach, the doctor suddenly stopped and had a piece of the broach handle in his hand.The piece came off the side of the broach handle, not the broach.He did not believe that there was a retractor pressing against the broach handle at the time of impaction.He removed the piece of metal, and continued to impact to the next size broach, using the same broach handle with no issues.He finished the case with the same broach handle and had no other issues.I am returning the broach handle for further inspection.The piece of the broach handle broke, fell into the patient, and then removed by the doctor by hand.We had to examine the piece to ensure it was from the broach handle and that everything was removed properly.Patient was last known to be in stable condition following the event.The device will be returned.
|
| |
|
Additional Manufacturer Narrative
|
|
Section h10: (h3) per the hhe, the most probable root cause of the reported event is design-related.The handle body subcomponent 301-300- 01 is a single body construction.However, the bridge feature that is fracturing is relatively thin (.035) and thus potentially susceptible to fracture.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional and/or corrected information.The following sections were corrected: h6.The complaint device was evaluated, and the reported event was confirmed.The evaluation identified.(high level observation that led them to associate the device with the capa: " the instrument was fracture").A definitive root cause was unable to be determined at this time; however, further investigation into the reported event is being conducted through our capa process.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional and/or corrected information.The complaint device was evaluated, and the reported event was confirmed.The evaluation identified.(high level observation that led them to associate the device with the capa: " the instrument was fracture").A definitive root cause was unable to be determined at this time; however, further investigation into the reported event is being conducted through our capa process.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|