| Model Number |
R SERIES |
| Medical Device Problem Code |
Unexpected Shutdown (4019)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during a routine shift check by a clinician, the device inappropriately shut down.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to a zoll medical corporation.The customer's report was not replicated or confirmed.The device was put through extensive testing including power cycling and full functionality testing using test battery and ac power cord without duplicating the report.Review of the device log shows that the device was prompting "low battery" and "replace battery" messages.The investigation concluded that the device met specifications and performed as designed.The device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
|
| |
|