TELEFLEX MEDICAL HUDSON NEBULIZER ADAPTOR 028,STERILE,SHELFPAK,; NEBULIZER (DIRECT PATIENT INTE Back to Search Results
Model Number IPN040618
Medical Device Problem Code Connection Problem (2900)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/02/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed and no relevant findings were identified.Without the device to evaluate, the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.If the sample becomes available at a later date a follow up report will be submitted with investigation results.
 
Event or Problem Description
It was reported that "the user found the connection between the adaptor and the flowmeter was unstable when starting to use.Therefore, the adaptor was replaced with a new unit".No patient harm reported.Patient condition reported as "fine".
 
Additional Manufacturer Narrative
(b)(4).The sample was returned for evaluation.A visual exam was performed and no defects were observed.Oxygen entrainment testing was performed and during the setup it was observed that the assembly of the nut adaptor and the upper body was unstable.Even with that condition, the sample was able to be tested on oxygen entrainment testing with no functional issues.After the testing finished, the adaptor was carefully disassembled from the upper body and it was visually inspected.During the visual inspection wear was found on the internal tabs.A device history record review was performed and no relevant findings were identified.Based on the investigation performed, the complaint is confirmed.Although the complaint is confirmed, there is not sufficient evidence to assure that this issue was originated during the manufacturing assembly or molding process.The wear on the internal tabs of the adaptor was most likely caused by the end user during the connection of the adaptor into the flowmeter that causes an unstable connection issue.
 
Event or Problem Description
It was reported that "the user found the connection between the adaptor and the flowmeter was unstable when starting to use.Therefore, the adaptor was replaced with a new unit".No patient harm reported.Patient condition reported as "fine".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUDSON NEBULIZER ADAPTOR 028,STERILE,SHELFPAK,
Common Device Name
NEBULIZER (DIRECT PATIENT INTE
Manufacturer (Section D)
TELEFLEX MEDICAL
morrisville NC
MDR Report Key11365250
Report Number3004365956-2021-00041
Device Sequence Number1101910
Product Code CAF
UDI-Device Identifier14026704596925
UDI-Public14026704596925
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K153010
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Expiration Date10/08/2024
Device Model NumberIPN040618
Device Catalogue Number031-28J
Device Lot Number74K1900596
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/24/2021
Initial Date Received by Manufacturer 02/03/2021
Supplement Date Received by Manufacturer03/15/2021
Initial Report FDA Received Date02/23/2021
Supplement Report FDA Received Date03/16/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-