• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NEUROMODULATION INTELLIS; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NEUROMODULATION INTELLIS; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF Back to Search Results
Model Number 97725
Medical Device Problem Code Positioning Problem (3009)
Health Effect - Clinical Codes Pain (1994); Burning Sensation (2146); Paresthesia (4421)
Date of Event 02/19/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Concomitant medical products: product id: 977d260, serial#: (b)(4), implanted: (b)(6) 2021, explanted: (b)(6) 2021, product type: screening device.Product id: 977d260, serial#: (b)(4), implanted: (b)(6) 2021, explanted:(b)(6) 2021, product type: screening device.Other relevant device(s) are: product id: 977d260, serial/lot #: (b)(4), ubd: 28-jan-2025, udi#: (b)(4); product id: 977d260, serial/lot #: (b)(4), ubd: 07-jan-2025, udi#: (b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that patient woke up from anesthesia with burning, tingling, and pain in both of her legs.This was prior to the device being turned on or tested.Rep reported difficulty placing the leads during the procedure.Attempted to test stimulator on (b)(6)2021, but did not change patient's pain.Leads were removed before leaving surgery center.Issue not resolved at time of report.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
The cause of the burning, tingling, and pain in both legs was not determined.The patient went to the hospital following this to have testing done.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLIS
Common Device Name
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF
Manufacturer (Section D)
MEDTRONIC NEUROMODULATION
7000 central ave ne
minneapolis MN 55432 1200
MDR Report Key11365979
Report Number3007566237-2021-00024
Device Sequence Number13714733
Product Code LGW
UDI-Device Identifier00763000258115
UDI-Public00763000258115
Combination Product (Y/N)N
PMA/510(K) Number
P840001
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,consum
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 03/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/26/2022
Device Model Number97725
Device Catalogue Number97725
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/19/2021
Supplement Date Received by Manufacturer03/04/2021
Initial Report FDA Received Date02/23/2021
Supplement Report FDA Received Date03/07/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age42 YR
-
-