Model Number ACU0T0 |
Device Problem
Activation, Positioning or Separation Problem (2906)
|
Patient Problem
Failure of Implant (1924)
|
Event Date 01/08/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Evaluation summary: complaint history records were reviewed.Product history records were reviewed and the documentation indicated the product met release criteria.The root cause has not been identified.There have been three other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
|
|
Event Description
|
A customer reported while using a preloaded intraocular lens (iol) delivery system, the plunger over-rode the lens.A backup lens was used to complete the procedure.Additional information was requested.
|
|
Manufacturer Narrative
|
Only the device was returned.The plunger lock and lens stop were removed.The plunger was oriented correctly.Viscoelastic was observed in the device.The plunger was retracted to mid-nozzle.No damage was observed to the device.The nozzle was removed and cleaned for further evaluation.The lens was removed from the nozzle during cleaning.Top coat dye stain testing was conducted with acceptable results.Product history records were reviewed and the documentation indicated the product met release criteria.A qualified viscoelastic was indicated.The root cause cannot be determined for the complaint of plunger override.Only the device was returned.It is not possible to confirm the position of the lens during delivery in relation to the plunger inside the device without the presence of the lens upon return.Top coat dye stain testing was conducted with acceptable results.The manufacturer internal reference number is: (b)(4).
|
|
Search Alerts/Recalls
|