MAUDE Adverse Event Report: IMPLANT DIRECT SYBRON MANUFACTURING LLC LEGACY 1 IMPLANT; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

IMPLANT DIRECT SYBRON MANUFACTURING LLC LEGACY 1 IMPLANT; DENTAL IMPLANT Back to Search Results
Catalog Number 805710
Medical Device Problem Code Loss of Osseointegration (2408)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 10/08/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Patient's identifier was not provided.When the requested information becomes available, a supplementary report will be submitted.Device evaluation results are not available.When the analysis is complete, a supplemental report will be submitted.
 
Event or Problem Description
Per complaint (b)(4), after clinical procedure, implant fracture was observed.
 
Additional Manufacturer Narrative
Follow-up submitted to report device evaluation.Updated section b4 for report submission date and b6 to report device evaluation results.Updated d4 for catalog # and unique identifier (udi) #, d9 for device return date, g1 for follow-up report submitter, g3 for awareness date and g6 for report type and follow-up number.Updated h2 for follow-up type, h3 for device evaluation status and h6 method, result and conclusion codes.Device received is different from the part that was indicated on the initial 3500a report.Lot number and manufacture/expiration dates previously submitted do not apply.Lot number information is unknown.Pma/510(k) number was updated.This complaint is being submitted late due to the furloughs that resulted from the global pandemic and is captured within deviation 1412.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEGACY 1 IMPLANT
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
IMPLANT DIRECT SYBRON MANUFACTURING LLC
3050 east hillcrest drive
thousand oaks CA 91362
MDR Report Key11366104
Report Number3001617766-2021-01457
Device Sequence Number19030794
Product Code DZE
UDI-Device Identifier10841307105862
UDI-Public10841307105862
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K061319
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Dentist
Remedial Action Inspection
Type of Report Initial,Followup
Report Date (Section B) 05/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/04/2013
Device Catalogue Number805710
Device Lot Number19880
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/16/2021
Initial Date Received by Manufacturer 02/10/2021
Supplement Date Received by Manufacturer03/22/2021
Initial Report FDA Received Date02/23/2021
Supplement Report FDA Received Date05/05/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age69 YR
Patient Weight73
-
-