| Model Number |
4468 |
| Medical Device Problem Code |
Material Puncture/Hole (1504)
|
| Health Effect - Clinical Codes |
No Known Impact Or Consequence To Patient (2692); Not Applicable (3189)
|
| Date of Event |
12/15/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Initial emdr submission.A follow up emdr will be submitted if additional information becomes available.(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the gloves has holes.
|
| |
|
Additional Manufacturer Narrative
|
|
A sample was available for evaluation.Visual examination was done according to our internal procedures and holes in the gloves were observed.As a result, bd was able to verify the reported issue.Material could be related to the failure as there is a small chance that an affected glove wallet was included in the involved batch.The supplier for this material was informed and asked to start an investigation to reach a root cause.The supplier of the gloves, baldur systems corp, was notified to trigger an investigation and reach a root cause.No further actions are required at this time.This failure will continue to be tracked and trended.H3 other text : see narrative below.
|
| |
|
Event or Problem Description
|
|
It was reported that the gloves has holes.
|
| |
|