Model Number 799-AEB1-00-01
Device Problem Communication or Transmission Problem (2896)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
 
Event Description
Complainant alleged that during incoming inspection, the device displayed an unknown "internal comm" error message.Complainant indicated that there was no patient involvement in the reported malfunction.
 
Manufacturer Narrative
This supplemental medwatch report is reporting the evaluation of the device.This supplemental medwatch report is also updating information submitted on the initial medwatch report.Please reference section b5.Evaluation results: the customer's report was observed during intial tesing of the device.However, the device was put through extensive testing without duplicating the report.A system interconnection cable was replaced and the calibration table was reloaded as a precaution.The device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
 
Event Description
Complainant alleged that during incoming inspection, the device displayed an "internal comm failure 1474" error message.Complainant indicated that there was no patient involvement in the reported malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNI-VENT, MODEL 731
Type of Device
VENTILATOR
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
MDR Report Key11366653
MDR Text Key234419639
Report Number1220908-2021-00481
Device Sequence Number1
Product Code CBK
UDI-Device Identifier00847946024192
UDI-Public00847946024192
Combination Product (y/n)N
PMA/PMN Number
K111473
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Type of Report Initial,Followup
Report Date 02/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/23/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number799-AEB1-00-01
Device Catalogue Number799-AEB1-00-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/12/2021
Was the Report Sent to FDA? No
Date Manufacturer Received02/02/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-