| Model Number |
N/A |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/18/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the instrument was discovered fractured during inventory.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned product exhibits signs of repeated use and has fractured on the medial side of the post feature.All pieces were returned.The device history record(s) were reviewed and no discrepancies relevant to the reported event were found.Medical records were not provided.A definitive root cause cannot be determined.Evaluation of the returned device identified the fracture was consistent with the tasp fractures analyzed in a zrm.The zrm identified that the common failure modes for the tasp devices include either bending overload or low cycle fatigue culminating in bending overload as evident by the presence of hackle marks, river lines and striations features on the fracture surface.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|